Résumé
A system of instrumentation for continuous measurement of gaseous metabolism and minute volume (V̇I) in the human newborn is described. O2 uptake and CO2 production are measured by open-circuit techniques utilizing a Servomex OA184 differential paramagnetic O2 analyzer and a Beckman LB-2 infrared CO2 analyzer. V̇I is measured with bias-flow pneumotachometry. Bench performance is described, methodological errors are defined, and clinical data are presented. The instrumentation is capable of safe, accurate, and continuous measurement of respiratory and metabolic variables in low-birth-weight infants.
Langue d'origine | English |
---|---|
Pages (de-à) | 1098-1103 |
Nombre de pages | 6 |
Journal | Unknown Journal |
Volume | 50 |
Numéro de publication | 5 |
DOI | |
Statut de publication | Published - 1981 |
Financement
Bailleurs de fonds | Numéro du bailleur de fonds |
---|---|
National Heart, Lung, and Blood Institute | P50HL014218 |
ASJC Scopus Subject Areas
- Physiology
- Endocrinology
Empreinte numérique
Plonger dans les sujets de recherche 'Continuous measurement of minute ventilation and gaseous metabolism of newborn infants'. Ensemble, ils forment une empreinte numérique unique.Citer
Schulze, K., Kairam, R., Stefanski, M., Sciacca, R., & James, L. S. (1981). Continuous measurement of minute ventilation and gaseous metabolism of newborn infants. Unknown Journal, 50(5), 1098-1103. https://doi.org/10.1152/jappl.1981.50.5.1098