Quantitative Ultrashort Echo Time Magnetic Resonance Imaging Evaluation of Postoperative Menisci: a Pilot Study

Darryl B. Sneag, Parina Shah, Matthew F. Koff, Wei Y. Lim, Scott A. Rodeo, Hollis G. Potter

Résultat de rechercheexamen par les pairs

24 Citations (Scopus)

Résumé

Background: Magnetic resonance imaging (MRI) visualization of meniscal signal is particularly challenging as the highly organized ultrastructure of meniscal fibrocartilage yields very short T2 values (∼6 ms) and a paucity of signal intensity during conventional image acquisition. Question/Purpose: The purpose of this study was to evaluate the feasibility of imaging postoperative menisci using an experimental, quantitative ultrashort echo time (UTE) MRI pulse sequence. This sequence acquires short echo images (echo time (TE) ∼0.3 ms) to produce multi-echo images for quantitative T2* calculations that provide an objective measure of collagen organization. Patients and Methods: MRI scans of the knee were acquired at 6- and 12-month intervals on a clinical 3.0 T scanner following meniscal surgery in eight patients (ages 13–41), four of whom underwent repair and the other four partial meniscectomy. Conventional MRI sequences were qualitatively evaluated for meniscal morphology and signal and correlated with quantitative UTE results. Results: A wide range of mean T2* values for both postsurgical groups was measured, and these values changed for each patient between the 6- and 12-month intervals. In many instances, the UTE sequence demonstrated quantitative differences between the two time intervals that were not detected with conventional sequences. Conclusions: This pilot study presents preliminary, observational data to be used as a baseline for future studies. Although the T2* values did not reveal a trend in either group or correlate with expected signal changes on conventional MRI, we speculate that the UTE sequence may detect ultrastructural alterations in meniscal composition that are otherwise not perceived with routine fast spin echo (FSE) sequences.

Langue d'origineEnglish
Pages (de-à)123-129
Nombre de pages7
JournalHSS Journal
Volume11
Numéro de publication2
DOI
Statut de publicationPublished - nov. 18 2015

Financement

Darryl B. Sneag, MD, reports institutional research funding from General Electric Healthcare during the conduct of the study. Matthew F. Koff, PhD, reports grant funding from NIH/NIAMS, outside the work. Scott A. Rodeo, MD, is a paid consultant for Smith and Nephew and has stocks/stock options from Cayenne Medical. Hollis G. Potter, MD, reports institutional funding from General Electric Healthcare during the conduct of study and grant funding from NIH/NIAMS, outside the work.

Bailleurs de fondsNuméro du bailleur de fonds
National Institute of Arthritis and Musculoskeletal and Skin Diseases
National Institutes of Health
National Institutes of Health
National Institute of Arthritis and Musculoskeletal and Skin Diseases

    ASJC Scopus Subject Areas

    • Surgery
    • Orthopedics and Sports Medicine

    Empreinte numérique

    Plonger dans les sujets de recherche 'Quantitative Ultrashort Echo Time Magnetic Resonance Imaging Evaluation of Postoperative Menisci: a Pilot Study'. Ensemble, ils forment une empreinte numérique unique.

    Citer