The effect of screw size and insertion technique on the stability of the mandibular sagittal split osteotomy

Alan Schwimmer, Alex M. Greenberg, Fredrick Kummer, Ayfer Kaynar

Producción científicarevisión exhaustiva

30 Citas (Scopus)

Resumen

The effect of screw size and insertion technique on the stability of sagittal split osteotomies was evaluated on 10 cadaver mandibles. The mandibles were osteotomized, and stabilized with either 2.0- or 2.7-mm stainless steel screws inserted as lag screws or position screws. All specimens were tested on an MTS servohydraulic testing machine (MTS Systems, Inc, Minneapolis, MN) to determine the strength and stability of the fixation. There were no statistically significant differences between the strengths achieved with the different screw sizes or fixation techniques. The primary variable that determined stability seemed to be the quality of the underlying bone.

Idioma originalEnglish
Páginas (desde-hasta)45-48
Número de páginas4
PublicaciónJournal of Oral and Maxillofacial Surgery
Volumen52
N.º1
DOI
EstadoPublished - ene. 1994

Financiación

FinanciadoresNúmero del financiador
ASIF
Swiss Association for Internal Fixation) Foundation

    ASJC Scopus Subject Areas

    • Surgery
    • Oral Surgery
    • Otorhinolaryngology

    Huella

    Profundice en los temas de investigación de 'The effect of screw size and insertion technique on the stability of the mandibular sagittal split osteotomy'. En conjunto forman una huella única.

    Citar esto